One-Pager

Hemodynamic Normals

Normal adult values, key formulas and how to read the waveforms behind them

Reviewed by Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P · Checked against current sources October 2026

Download one-page PDFLetter landscape, ready to print or keep with your notes

Pressure normals

Typical adult ranges.
ParameterNormalUnits
HR60–100beats/min
MAP70–105mmHg
CVP / RAP2–6mmHg
RV15–25 / 0–8mmHg
PA15–25 / 8–15mmHg
PA mean10–20mmHg
PAOP (wedge)6–12mmHg

Preload: CVP (right), PAOP (left). Afterload: PVR (right), SVR (left). Pulmonary hypertension: mean PA above 20 mmHg since WSPH 2018; older texts use 25 or more.

Flow, resistance and oxygen

1 WU = 80 dyn·s/cm⁵; PH guidelines call PVR above 2 WU (160) high.
ParameterNormalUnits
CO4–8L/min
CI2.5–4.0L/min/m²
SV60–100mL/beat
SVI33–47mL/m²/beat
SVR800–1200dyn·s/cm⁵
SVRI1970–2390dyn·s/cm⁵·m²
PVR100–250dyn·s/cm⁵
SvO260–80%
ScvO2About 70%
DO2950–1150mL/min

Arterial line and square-wave test

04080120160mmHg
Normal: brisk upstroke, systolic peak, dicrotic notch (aortic valve closure).

Square-wave test: snap the fast flush and count oscillations before the tracing returns.

ResultOscillationsError
Optimal1.5–2None
OverdampedFewer than 1.5False low SBP, high DBP
UnderdampedMore than 2False high SBP, low DBP
04080120160mmHg
Overdamped: slurred upstroke, no notch, narrow pulse pressure. Look for air, clots, kinks or a soft flush bag (keep 300 mmHg).
04080120160mmHg
Underdamped: overshoot spike with ringing; systolic reads falsely high. Check for long or added tubing.

CVP and PA catheter waveforms

05101520mmHg
CVP: a wave after P (atrial kick), c after QRS (tricuspid closure), v just after T (atrial filling).
010203040mmHg
Insertion: RA, RV (diastolic near 0), PA (diastolic step-up, dicrotic notch), then wedge.
010203040mmHg
Balloon up: PA trace drops to wedge, mean below PA diastolic.

Reading the numbers

Fluid responsiveness

Practice Hemodynamic Normals questions free in your browser5 exam-style questions with rationales. No account or download needed.

In depth

The one-page summary above is what the PDF prints. These sections go further.

Cardiac output and its four determinants

Cardiac output is heart rate times stroke volume, and stroke volume is set by preload, afterload and contractility. Output carries oxygen: DO₂ = CO × CaO₂ × 10, so a CO of 5 L/min with a CaO₂ of 20 mL/dL delivers 1,000 mL/min against a resting VO₂ of about 200–250 mL/min, an extraction near 25%. Every monitor on this page measures one of these pieces or a pressure that stands in for one.

DeterminantBedside estimateRaised byLowered by
PreloadCVP, PAOP, PPV/SVV, echoFluid, leg raise, venoconstrictionBleeding, diuresis, venodilators, PEEP
AfterloadSVR, PVR, MAPVasopressors, hypertension, ASVasodilators, sepsis, IABP
ContractilityEF, upstroke slope, CIInotropes, catecholaminesIschemia, acidosis, sepsis, beta-blockers
Heart rate and rhythmECGPacing, chronotropesBeta-blockers, block, loss of atrial kick

Frank-Starling: on the steep part of the curve more end-diastolic volume gives more stroke volume; on the flat part it only raises filling pressures and drives edema. A failing ventricle has a flatter curve, and inotropes lift it. CVP and PAOP stand in for volume only through compliance, so a stiff ventricle (hypertrophy, ischemia, tamponade, high PEEP) shows a high pressure with a small volume.

Afterload is wall stress during ejection (Laplace: pressure × radius ÷ 2 × wall thickness), so a dilated ventricle faces more of it at the same blood pressure; SVR is only a calculated surrogate. A failing LV loses stroke volume steeply as SVR rises, which is why afterload reduction and IABP raise output in cardiogenic shock. The thin RV is even more afterload-sensitive: a previously normal RV cannot generate a mean PA pressure much above 40 mmHg, so massive PE presents as RV failure rather than high PA pressure. Rate raises output until diastole is too short to fill, often from about 120–150 in adults and earlier with a stiff ventricle; the atrial kick supplies about 15–30% of LV filling, so atrial fibrillation or VVI pacing can drop output with no change in volume.

Transducer physics: leveling, zeroing, dynamic response

Zeroing opens the transducer to air so the monitor subtracts atmospheric pressure; do it at setup, after disconnection and whenever the numbers do not fit the patient. Leveling puts the air-fluid interface at the phlebostatic axis. Each 1.36 cm of height error is 1 mmHg, so a 10 cm error is about 7 mmHg: trivial against a MAP of 80, but it turns a CVP of 8 into 15 or 1. Re-level pole-mounted transducers after every change in bed height; HOB 0–60° supine is acceptable, and lateral positions need the angle-specific reference points in the AACN 2016 alert.

Dynamic response: catheter, tubing and diaphragm form a spring-mass system with a natural frequency (how fast it rings) and a damping coefficient (how fast the ringing dies). The arterial pulse contains harmonics up to about ten times the heart rate, so a system with too low a natural frequency resonates and overshoots systolic. Bedside kits sit around 10–20 Hz, so damping must be near the optimum of about 0.6–0.7 (Gardner 1981). On a printed square-wave test, natural frequency is paper speed divided by the distance between oscillation peaks (25 mm/s ÷ 1.5 mm ≈ 17 Hz); at the bedside, 1.5–2 oscillations is the practical proxy.

ProblemCommon causesFix
OverdampedAir, clot at tip, kink, soft bag, flexed wristRemove air, aspirate and flush, bag to 300
UnderdampedLong or added tubing, tachycardia, stiff arteriesShorten tubing, damping device, use MAP
PA catheter whipTip moving in a hyperdynamic PARead the mean; consider repositioning

Arterial waveform analysis

The trace has an upstroke (its slope reflects contractility), a systolic peak, a dicrotic notch (aortic valve closure plus reflected waves) and diastolic runoff; the area under the systolic portion tracks stroke volume, which pulse-contour monitors exploit. From aorta to radial artery, reflected waves raise the systolic peak (often 10–20 mmHg) and lower the diastolic slightly, while MAP falls only 1–2 mmHg; hence vasopressors are titrated to MAP. In septic shock on high-dose vasopressors, after bypass and in hypothermia, radial MAP can read several mmHg below central pressure (Dorman 1998), so consider a femoral line when the radial number does not fit the patient.

PatternWhat you seeThink of
Pulsus paradoxusSBP falls more than 10 mmHg on inspirationTamponade, severe asthma, massive PE
Pulsus alternansStrong and weak beats alternateSevere LV failure
Parvus et tardusSlow upstroke, late low peakAortic stenosis (or overdamping)
BisferiensTwo systolic peaksAortic regurgitation, HOCM
Wide pulse pressure, low notchLow diastolic, notch near baselineVasodilation, sepsis, AR
Narrow pulse pressureSmall beatsHypovolemia, tamponade, cardiogenic shock

On positive-pressure ventilation, inspiration first squeezes blood from the lungs into the LV (systolic rises), then cuts RV filling so LV output falls a few beats later. A large fall means preload dependence and is the basis of pulse pressure variation.

CVP: what it measures and where it fails

CVP is right atrial pressure and, without tricuspid stenosis, RV end-diastolic pressure. It is also the back pressure for venous return (Guyton: mean systemic filling pressure minus RAP, divided by venous resistance), so a high CVP obstructs flow and lowers organ perfusion pressure; kidney perfusion is roughly MAP − CVP.

As a guide to fluid, CVP is a coin toss in the normal range: an area under the curve of 0.56 in the Marik 2013 meta-analysis. Extremes help a little: under about 6–8 makes response more likely, above about 12–15 less likely (De Backer and Vincent 2018). Its best use is as a safety limit: if CVP climbs with a bolus while stroke volume does not, the RV is on the flat part of its curve. Besides volume, CVP rises with RV failure, pulmonary hypertension, PE, tricuspid regurgitation, tamponade, PEEP and auto-PEEP, abdominal hypertension and tension pneumothorax.

FindingMechanismCauses
Large a waveAtrium contracts against resistanceRV hypertrophy, pulmonary hypertension, TS
Cannon a waveAtrium contracts against a closed TVAV dissociation, VVI pacing, VT
No a waveNo organized atrial contractionAtrial fibrillation or flutter
Large v (cv) waveBackflow during systoleTricuspid regurgitation
Steep y, M or W shapeRapid early filling, then abrupt haltConstrictive pericarditis
Blunted y, CVP ≈ PAD ≈ PAOPDiastolic equalizationTamponade

With no a wave (atrial fibrillation), read CVP at the z point, at the end of the QRS just before the c wave. With large v waves, the digital mean overestimates filling pressure.

PA catheter: insertion, waveforms and wedge validity

The adult catheter is 110 cm, 7–8 Fr, with marks every 10 cm. The distal lumen reads PA pressure and gives mixed venous samples; the proximal lumen, about 30 cm back, sits in the RA for CVP and injectate; the balloon takes 1.5 mL; a thermistor near the tip reads blood temperature. From the right internal jugular the RA is reached at about 15–20 cm and wedge usually at 45–55 cm; subclavian is about 5 cm shorter, femoral about 10–15 cm longer (Edwards Quick Guide).

The balloon is inflated in the RA and the catheter floats with the flow. RA shows a and v waves; RV has a diastolic near 0 that rises through diastole; PA has a diastolic step-up that falls through diastole and a dicrotic notch; wedge is damped with a and v waves. No RV trace within about 20 cm of the RA suggests coiling: deflate, withdraw to the RA and retry. Withdraw only with the balloon down. Watch for ventricular ectopy as the tip crosses the RV and for right bundle branch block, which with existing LBBB means complete heart block, so have pacing available. Final position: nearly the full 1.25–1.5 mL is needed to wedge and the PA trace returns at once on deflation.

PAOP reflects left atrial pressure only through a continuous column of blood to the LA, which requires West zone 3, where pulmonary arterial and venous pressures exceed alveolar pressure. In zones 1 and 2 the catheter reads alveolar pressure instead. Hypovolemia and high PEEP enlarge those zones; flow-directed catheters usually float to dependent zone 3. Signs the wedge is valid:

Read at end-expiration, when pleural pressure is closest to atmospheric: just before the inspiratory dip when breathing spontaneously, just before the inspiratory rise on the ventilator. The digital display can be wrong by the full respiratory swing, so use a frozen trace beside the airway pressure or capnography waveform. Active exhalation raises end-expiratory values falsely and has no clean fix. PEEP above about 10 cmH₂O makes PAOP overestimate transmural LA pressure (Bootsma 2022), but never take the patient off PEEP to measure: derecruitment is harmful and the number no longer describes the patient as treated.

Measuring flow and oxygen: thermodilution, Fick, SvO₂

Bolus thermodilution: cold injectate enters the RA through the proximal port and the thermistor records the temperature curve in the PA; output is inversely proportional to the area under the curve (Stewart-Hamilton). Inject 10 mL at least 10 °C colder than blood smoothly in under 4 seconds at end-expiration; average three values within about 10% of each other. Continuous-output catheters heat blood with a filament and report a rolling average over several minutes, so they lag acute changes.

ErrorEffect on CO reading
Less injectate than the computer expectsFalsely high
Injectate warmed in the handFalsely high
Slow or uneven injectionFalsely low
Tricuspid regurgitationUnreliable, usually falsely low
Left-to-right shunt (VSD)Measures pulmonary flow, overstates systemic

Fick: CO = VO₂ ÷ (CaO₂ − CvO₂) × 10. VO₂ is rarely measured outside the cath lab, and the assumed 125 mL/min/m² can be badly wrong in fever, sepsis or sedation, so indirect Fick is a cross-check. The normal arteriovenous content difference is about 4–6 mL/dL and widens as output falls.

FeatureSvO₂ScvO₂
Sample sitePA, distal portSVC, CVC tip
Normal60–80%About 70%
Blood includedSVC, IVC and coronary sinusUpper body only
In healthSlightly higherAbout 2–3% lower
In shockLowerAbout 7% higher, gap widens

Low SvO₂: delivery is not meeting demand (low output, hemoglobin or SaO₂, or fever, shivering, seizures, pain, work of breathing). High SvO₂ above 80%: tissues are not extracting (sepsis, cyanide from nitroprusside, hypothermia, left-to-right shunt) or a wedged catheter is sampling arterialized blood. A normal ScvO₂ does not exclude hypoperfusion; a venous-to-arterial PCO₂ gap above 6 mmHg suggests flow is still inadequate. ProCESS 2014, ARISE 2014 and ProMISe 2015 found no mortality benefit from ScvO₂-targeted early goal-directed therapy over usual care.

Derived values and worked profiles

ValueFormulaNormal or threshold
SVR(MAP − CVP) × 80 ÷ CO800–1200 dyn·s/cm⁵
PVR(mean PA − PAOP) × 80 ÷ CO100–250; above 2 WU is high
Diastolic pulmonary gradientPAD − PAOPUnder 7 mmHg
Cardiac power output (CPO)MAP × CO ÷ 451About 1 W at rest; under 0.6 W poor
PAPi(PA systolic − PA diastolic) ÷ RAPLow values flag RV failure
RAP:PAOP ratioRAP ÷ PAOPAbove about 0.6 suggests RV failure

CPO is the heart's hydraulic work. In the SHOCK trial registry it was the strongest hemodynamic predictor of in-hospital death, with a cutoff of about 0.53 W (Fincke 2004). PAPi asks how well the RV turns filling pressure into pulsatile PA pressure: under 0.9 predicted RV failure or death in inferior MI (Korabathina 2012), and under about 1.85 predicted RV failure after LVAD (Morine 2016). Thresholds vary by population; treat them as flags.

Worked example: BP 84/52 (MAP 63), HR 110, CVP 14, PA 48/28 (mean 35), PAOP 26, CO 3.2 L/min, BSA 2.0. CI = 1.6 and SV ≈ 29 mL. SVR = (63 − 14) × 80 ÷ 3.2 ≈ 1,225. PVR = (35 − 26) × 80 ÷ 3.2 = 225 (2.8 WU) with a diastolic gradient of 2, so the pulmonary hypertension is post-capillary. CPO = 63 × 3.2 ÷ 451 ≈ 0.45 W; PAPi = 20 ÷ 14 ≈ 1.4; RAP:PAOP ≈ 0.54. Reading: LV-predominant cardiogenic shock with inadequate power and an RV still coping, the profile in which inotropes or LV mechanical support are considered.

Illustrative numbers showing direction, not cutoffs.
ProfileCVPPAOPCISVRSvO₂
Hypovolemic141.81,60055
Cardiogenic (LV)12261.61,50048
Septic, after fluid694.550080
Massive PE, RV failure1881.71,50050
Tamponade20201.61,50050
Neurogenic362.660072

Minimally invasive monitors, echo and fluid tests

MonitorHow it worksMain limits
Uncalibrated pulse contourArterial waveform plus demographic modelChanging tone, high-dose pressors, AR, IABP
Calibrated pulse contourTranspulmonary thermodilution resets itNeeds CVC and thermistor arterial line
Bioreactance (NICOM)Phase shift of current across the thoraxElectrodes, motion; mixed agreement
Esophageal DopplerDescending aortic velocity at T5–T6Sedated, intubated; probe drifts

These devices follow trends better than absolute values; uncalibrated pulse contour often misses the 30% percentage-error standard for agreement with thermodilution (Critchley 1999) in vasoplegia and arrhythmia. PPV above 13% predicts response (Michard 2000), but 9–13% is a gray zone holding about a quarter of patients (Cannesson 2011). PPV and SVV also need a heart rate to respiratory rate ratio above about 3.6 and no RV failure, beyond the conditions on the one-pager, and point-prevalence studies (Mahjoub 2014, Delannoy 2016) found few ICU patients meeting all of them. Under 6 mL/kg ventilation, a tidal volume challenge restores some value.

Echo questions a nurse can follow: is the LV squeezing well, poorly or hyperdynamically; is the RV larger than the LV or the septum D-shaped; is there pericardial fluid with RA or RV diastolic collapse? IVC under 2.1 cm collapsing more than 50% on a sniff suggests RAP about 3, over 2.1 cm collapsing less suggests about 15 (ASE 2015); the IVC is unreliable on the ventilator. LVOT VTI is normally about 18–22 cm.

TestHowPositiveLimits
Passive leg raiseFrom 45° semi-recumbent, trunk flat, legs upCO or SV up 10% within 1–2 minNeeds real-time CO; IAH; high ICP
Mini-fluid challenge100 mL over 1 min (Muller 2011)VTI up 10%Near the noise of echo
End-expiratory occlusionVentilator hold at end-expiration, 15 sCO up 5% (Monnet 2009)No triggering; needs pulse contour
Tidal volume challengeVT 6 to 8 mL/kg for 1 minPPV up 3.5 points (Myatra 2017)Controlled ventilation, regular rhythm
Fluid challenge250–500 mL over 10–15 minSV up 10–15%Gives the fluid whatever the answer

Only about half of ICU patients given fluid raise their output (Michard and Teboul 2002). The leg raise moves about 300 mL of venous blood into the chest and is reversible; the Monnet 2016 meta-analysis gave sensitivity 85% and specificity 91% when judged by output, and much worse by pulse pressure. Pain, coughing or waking during it give false positives. Each test's threshold must exceed its monitor's noise: a 5% rise shows on continuous pulse contour but not reliably on echo. Responsiveness is not an indication; also ask whether low flow is the problem and whether the patient tolerates more volume.

PA catheter complications and evidence

TrialPopulationResult
Connors 19965,735 ICU patients, observationalPAC linked to higher 30-day mortality
Sandham 20031,994 high-risk surgical patientsNo mortality difference; more PE with PAC
PAC-Man 20051,041 UK ICU patientsHospital mortality 68% vs 66%
ESCAPE 2005433 decompensated HFDays alive out of hospital 133 vs 135
FACTT 20061,000 ALI/ARDS patientsNo mortality difference; more catheter problems
Garan 2020 (CSWG)Cardiogenic shock registryEarly full profiling linked to lower mortality

Routine PAC placement does not improve outcomes in general ICU, surgical, ARDS or heart failure patients, and ESCAPE recorded more in-hospital adverse events with it (about 22% vs 12%). It is a monitor, so outcome depends on reading it well; Iberti 1990 found nearly half of surveyed physicians could not identify a wedge on a clear tracing. Use has narrowed to cardiogenic shock (staging, choosing and weaning mechanical support, judging the RV), pulmonary hypertension, transplant and LVAD care, and complex cardiac surgery. Whether PAC-guided care improves survival in cardiogenic shock is unsettled: the supporting data are observational.

Exam pearls

Practice free in your browser2,450+ exam-style questions with rationales. No account, no ads. Also free on the App Store and Google Play.

Related one-pagers

Calculators for this topic

Sources

  1. Edwards Lifesciences. Normal Hemodynamic Parameters and Laboratory Values (Edwards Clinical Education pocket card, PP--US-2312 v2.0). 2022. Accessed 2026-10; the file is marked as expiring 2026-04. Link
  2. Edwards Lifesciences. Normal Hemodynamic Parameters and Laboratory Values (earlier pocket card). 2014. Link
  3. Edwards Clinical Education. Quick Guide to Cardiopulmonary Care, 4th ed. Edwards Lifesciences. Link
  4. AACN Practice Alert: Pulmonary Artery/Central Venous Pressure Monitoring in Adults. Crit Care Nurse. 2016;36(4):e12-e18 (reviewed 2021). Link
  5. Bootsma IT, Boerma EC, de Lange F, Scheeren TWL. The contemporary pulmonary artery catheter. Part 1: placement and waveform analysis. J Clin Monit Comput. 2022;36(1):5-15. Link
  6. Bootsma IT, Boerma EC, Scheeren TWL, de Lange F. The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications. J Clin Monit Comput. 2022. Link
  7. Kovacs G, Bartolome S, Denton CP, et al. Definition, classification and diagnosis of pulmonary hypertension (7th World Symposium on Pulmonary Hypertension). Eur Respir J. 2024;64(4):2401324. doi:10.1183/13993003.01324-2024 Link
  8. Simonneau G, Montani D, Celermajer DS, et al. Haemodynamic definitions and updated clinical classification of pulmonary hypertension (6th WSPH, 2018). Eur Respir J. 2019;53(1):1801913. Link
  9. CHEST Physician. Major takeaways from the seventh World Symposium on PH. Link
  10. Prescott HC, Antonelli M, Alhazzani W, et al. Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock 2026. Crit Care Med. 2026. doi:10.1097/CCM.0000000000007075 Link
  11. Azevedo LCP, Myatra SN, Hammond N. Using the 2026 Surviving Sepsis Campaign Guidelines in Practice. Crit Care Sci. 2026;38:e20260036. Link
  12. Michard F, Boussat S, Chemla D, et al. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med. 2000;162(1):134-138. Link
  13. Monnet X, Marik P, Teboul JL. Passive leg raising for predicting fluid responsiveness: a systematic review and meta-analysis. Intensive Care Med. 2016;42(12):1935-1947. Link
  14. Monnet X, Teboul JL. Passive leg raising: five rules, not a drop of fluid! Crit Care. 2015;19:18. Link
  15. Delannoy B, Wallet F, Maucort-Boulch D, et al. Applicability of pulse pressure variation during unstable hemodynamic events in the intensive care unit: a five-day prospective multicenter study. Crit Care Res Pract. 2016. Link
  16. Pearse RM, Hinds CJ. Should we use central venous saturation to guide management in high-risk surgical patients? Crit Care. 2006;10(6):181. Link
  17. Brittain EL, et al. Predictors of diastolic-to-wedge gradient in patients evaluated for pulmonary hypertension. PLoS One. 2013;8(10):e76461. Link
  18. Deranged Physiology. Arterial line dynamic response testing (secondary educational source). Link
  19. Deranged Physiology. Arterial line pressure transducer setup (secondary educational source). Link
  20. Zhao F, Wei H, Lin L, et al. Levosimendan versus dobutamine in septic cardiomyopathy: a randomized clinical trial. Front Cardiovasc Med. 2025. Link