One-Pager
Hemodynamic Normals
Normal adult values, key formulas and how to read the waveforms behind them
Download one-page PDFLetter landscape, ready to print or keep with your notes
Pressure normals
| Parameter | Normal | Units |
|---|---|---|
| HR | 60–100 | beats/min |
| MAP | 70–105 | mmHg |
| CVP / RAP | 2–6 | mmHg |
| RV | 15–25 / 0–8 | mmHg |
| PA | 15–25 / 8–15 | mmHg |
| PA mean | 10–20 | mmHg |
| PAOP (wedge) | 6–12 | mmHg |
Preload: CVP (right), PAOP (left). Afterload: PVR (right), SVR (left). Pulmonary hypertension: mean PA above 20 mmHg since WSPH 2018; older texts use 25 or more.
Flow, resistance and oxygen
| Parameter | Normal | Units |
|---|---|---|
| CO | 4–8 | L/min |
| CI | 2.5–4.0 | L/min/m² |
| SV | 60–100 | mL/beat |
| SVI | 33–47 | mL/m²/beat |
| SVR | 800–1200 | dyn·s/cm⁵ |
| SVRI | 1970–2390 | dyn·s/cm⁵·m² |
| PVR | 100–250 | dyn·s/cm⁵ |
| SvO2 | 60–80 | % |
| ScvO2 | About 70 | % |
| DO2 | 950–1150 | mL/min |
- MAP = (SBP + 2 × DBP) ÷ 3
- CO = HR × SV ÷ 1000 (L/min); CI = CO ÷ BSA
- SVR = (MAP − CVP) × 80 ÷ CO
- PVR = (mean PA − PAOP) × 80 ÷ CO
- CaO2 = 1.34 × Hb × SaO2 (decimal) + 0.003 × PaO2 (17–20 mL/dL)
- DO2 = CO × CaO2 × 10
- Coronary perfusion = DBP − PAOP (60–80 mmHg)
Arterial line and square-wave test
Square-wave test: snap the fast flush and count oscillations before the tracing returns.
| Result | Oscillations | Error |
|---|---|---|
| Optimal | 1.5–2 | None |
| Overdamped | Fewer than 1.5 | False low SBP, high DBP |
| Underdamped | More than 2 | False high SBP, low DBP |
CVP and PA catheter waveforms
- PAOP estimates LA pressure and LVEDP; it reads high with mitral disease or high PEEP, low with a stiff LV.
- PAOP sits a few mmHg below PAD; a wide gap suggests high PVR. Small gap, stable patient: trend PAD and wedge less.
- Cannon a waves: AV dissociation. No a waves: atrial fibrillation.
- Giant v waves: tricuspid (CVP) or mitral (wedge) regurgitation.
Reading the numbers
- Level the air-fluid interface to the phlebostatic axis (4th intercostal space, midway front to back); re-level when position changes.
- Zero to air. Head of bed 0–60° is fine once leveled.
- Read at end-expiration on a frozen or printed trace, not the digital number: just before the inspiratory dip (spontaneous) or rise (ventilator).
- Read CVP and PAOP at the mean of the a wave.
Fluid responsiveness
- Responder: SV or CO rises more than 10% with a fluid challenge or passive leg raise (PLR).
- PLR: from 45° semi-recumbent, tilt the bed to trunk flat, legs up. Judge by SV or CO within 1–2 minutes, not BP.
- PPV above 13% or SVV above about 12% predicts response.
- PPV and SVV need controlled ventilation, no spontaneous breaths, regular rhythm, tidal volume about 8 mL/kg IBW, closed chest, normal abdominal pressure.
- SSC 2026: dynamic measures over physical exam or static values alone. Septic shock: initial MAP 65 mmHg (60–65 if 65 or older).
In depth
The one-page summary above is what the PDF prints. These sections go further.
Cardiac output and its four determinants
Cardiac output is heart rate times stroke volume, and stroke volume is set by preload, afterload and contractility. Output carries oxygen: DO₂ = CO × CaO₂ × 10, so a CO of 5 L/min with a CaO₂ of 20 mL/dL delivers 1,000 mL/min against a resting VO₂ of about 200–250 mL/min, an extraction near 25%. Every monitor on this page measures one of these pieces or a pressure that stands in for one.
| Determinant | Bedside estimate | Raised by | Lowered by |
|---|---|---|---|
| Preload | CVP, PAOP, PPV/SVV, echo | Fluid, leg raise, venoconstriction | Bleeding, diuresis, venodilators, PEEP |
| Afterload | SVR, PVR, MAP | Vasopressors, hypertension, AS | Vasodilators, sepsis, IABP |
| Contractility | EF, upstroke slope, CI | Inotropes, catecholamines | Ischemia, acidosis, sepsis, beta-blockers |
| Heart rate and rhythm | ECG | Pacing, chronotropes | Beta-blockers, block, loss of atrial kick |
Frank-Starling: on the steep part of the curve more end-diastolic volume gives more stroke volume; on the flat part it only raises filling pressures and drives edema. A failing ventricle has a flatter curve, and inotropes lift it. CVP and PAOP stand in for volume only through compliance, so a stiff ventricle (hypertrophy, ischemia, tamponade, high PEEP) shows a high pressure with a small volume.
Afterload is wall stress during ejection (Laplace: pressure × radius ÷ 2 × wall thickness), so a dilated ventricle faces more of it at the same blood pressure; SVR is only a calculated surrogate. A failing LV loses stroke volume steeply as SVR rises, which is why afterload reduction and IABP raise output in cardiogenic shock. The thin RV is even more afterload-sensitive: a previously normal RV cannot generate a mean PA pressure much above 40 mmHg, so massive PE presents as RV failure rather than high PA pressure. Rate raises output until diastole is too short to fill, often from about 120–150 in adults and earlier with a stiff ventricle; the atrial kick supplies about 15–30% of LV filling, so atrial fibrillation or VVI pacing can drop output with no change in volume.
Transducer physics: leveling, zeroing, dynamic response
Zeroing opens the transducer to air so the monitor subtracts atmospheric pressure; do it at setup, after disconnection and whenever the numbers do not fit the patient. Leveling puts the air-fluid interface at the phlebostatic axis. Each 1.36 cm of height error is 1 mmHg, so a 10 cm error is about 7 mmHg: trivial against a MAP of 80, but it turns a CVP of 8 into 15 or 1. Re-level pole-mounted transducers after every change in bed height; HOB 0–60° supine is acceptable, and lateral positions need the angle-specific reference points in the AACN 2016 alert.
Dynamic response: catheter, tubing and diaphragm form a spring-mass system with a natural frequency (how fast it rings) and a damping coefficient (how fast the ringing dies). The arterial pulse contains harmonics up to about ten times the heart rate, so a system with too low a natural frequency resonates and overshoots systolic. Bedside kits sit around 10–20 Hz, so damping must be near the optimum of about 0.6–0.7 (Gardner 1981). On a printed square-wave test, natural frequency is paper speed divided by the distance between oscillation peaks (25 mm/s ÷ 1.5 mm ≈ 17 Hz); at the bedside, 1.5–2 oscillations is the practical proxy.
| Problem | Common causes | Fix |
|---|---|---|
| Overdamped | Air, clot at tip, kink, soft bag, flexed wrist | Remove air, aspirate and flush, bag to 300 |
| Underdamped | Long or added tubing, tachycardia, stiff arteries | Shorten tubing, damping device, use MAP |
| PA catheter whip | Tip moving in a hyperdynamic PA | Read the mean; consider repositioning |
Arterial waveform analysis
The trace has an upstroke (its slope reflects contractility), a systolic peak, a dicrotic notch (aortic valve closure plus reflected waves) and diastolic runoff; the area under the systolic portion tracks stroke volume, which pulse-contour monitors exploit. From aorta to radial artery, reflected waves raise the systolic peak (often 10–20 mmHg) and lower the diastolic slightly, while MAP falls only 1–2 mmHg; hence vasopressors are titrated to MAP. In septic shock on high-dose vasopressors, after bypass and in hypothermia, radial MAP can read several mmHg below central pressure (Dorman 1998), so consider a femoral line when the radial number does not fit the patient.
| Pattern | What you see | Think of |
|---|---|---|
| Pulsus paradoxus | SBP falls more than 10 mmHg on inspiration | Tamponade, severe asthma, massive PE |
| Pulsus alternans | Strong and weak beats alternate | Severe LV failure |
| Parvus et tardus | Slow upstroke, late low peak | Aortic stenosis (or overdamping) |
| Bisferiens | Two systolic peaks | Aortic regurgitation, HOCM |
| Wide pulse pressure, low notch | Low diastolic, notch near baseline | Vasodilation, sepsis, AR |
| Narrow pulse pressure | Small beats | Hypovolemia, tamponade, cardiogenic shock |
On positive-pressure ventilation, inspiration first squeezes blood from the lungs into the LV (systolic rises), then cuts RV filling so LV output falls a few beats later. A large fall means preload dependence and is the basis of pulse pressure variation.
CVP: what it measures and where it fails
CVP is right atrial pressure and, without tricuspid stenosis, RV end-diastolic pressure. It is also the back pressure for venous return (Guyton: mean systemic filling pressure minus RAP, divided by venous resistance), so a high CVP obstructs flow and lowers organ perfusion pressure; kidney perfusion is roughly MAP − CVP.
As a guide to fluid, CVP is a coin toss in the normal range: an area under the curve of 0.56 in the Marik 2013 meta-analysis. Extremes help a little: under about 6–8 makes response more likely, above about 12–15 less likely (De Backer and Vincent 2018). Its best use is as a safety limit: if CVP climbs with a bolus while stroke volume does not, the RV is on the flat part of its curve. Besides volume, CVP rises with RV failure, pulmonary hypertension, PE, tricuspid regurgitation, tamponade, PEEP and auto-PEEP, abdominal hypertension and tension pneumothorax.
| Finding | Mechanism | Causes |
|---|---|---|
| Large a wave | Atrium contracts against resistance | RV hypertrophy, pulmonary hypertension, TS |
| Cannon a wave | Atrium contracts against a closed TV | AV dissociation, VVI pacing, VT |
| No a wave | No organized atrial contraction | Atrial fibrillation or flutter |
| Large v (cv) wave | Backflow during systole | Tricuspid regurgitation |
| Steep y, M or W shape | Rapid early filling, then abrupt halt | Constrictive pericarditis |
| Blunted y, CVP ≈ PAD ≈ PAOP | Diastolic equalization | Tamponade |
With no a wave (atrial fibrillation), read CVP at the z point, at the end of the QRS just before the c wave. With large v waves, the digital mean overestimates filling pressure.
PA catheter: insertion, waveforms and wedge validity
The adult catheter is 110 cm, 7–8 Fr, with marks every 10 cm. The distal lumen reads PA pressure and gives mixed venous samples; the proximal lumen, about 30 cm back, sits in the RA for CVP and injectate; the balloon takes 1.5 mL; a thermistor near the tip reads blood temperature. From the right internal jugular the RA is reached at about 15–20 cm and wedge usually at 45–55 cm; subclavian is about 5 cm shorter, femoral about 10–15 cm longer (Edwards Quick Guide).
The balloon is inflated in the RA and the catheter floats with the flow. RA shows a and v waves; RV has a diastolic near 0 that rises through diastole; PA has a diastolic step-up that falls through diastole and a dicrotic notch; wedge is damped with a and v waves. No RV trace within about 20 cm of the RA suggests coiling: deflate, withdraw to the RA and retry. Withdraw only with the balloon down. Watch for ventricular ectopy as the tip crosses the RV and for right bundle branch block, which with existing LBBB means complete heart block, so have pacing available. Final position: nearly the full 1.25–1.5 mL is needed to wedge and the PA trace returns at once on deflation.
PAOP reflects left atrial pressure only through a continuous column of blood to the LA, which requires West zone 3, where pulmonary arterial and venous pressures exceed alveolar pressure. In zones 1 and 2 the catheter reads alveolar pressure instead. Hypovolemia and high PEEP enlarge those zones; flow-directed catheters usually float to dependent zone 3. Signs the wedge is valid:
- Tip at or below the LA on a lateral film.
- Clear a and v waves; a smooth, flat trace suggests zone 1 or 2.
- PAOP lower than PAD.
- Respiratory swings in PAOP no larger than those in PAD.
- PAOP rises by less than half of any increase in PEEP.
Read at end-expiration, when pleural pressure is closest to atmospheric: just before the inspiratory dip when breathing spontaneously, just before the inspiratory rise on the ventilator. The digital display can be wrong by the full respiratory swing, so use a frozen trace beside the airway pressure or capnography waveform. Active exhalation raises end-expiratory values falsely and has no clean fix. PEEP above about 10 cmH₂O makes PAOP overestimate transmural LA pressure (Bootsma 2022), but never take the patient off PEEP to measure: derecruitment is harmful and the number no longer describes the patient as treated.
Measuring flow and oxygen: thermodilution, Fick, SvO₂
Bolus thermodilution: cold injectate enters the RA through the proximal port and the thermistor records the temperature curve in the PA; output is inversely proportional to the area under the curve (Stewart-Hamilton). Inject 10 mL at least 10 °C colder than blood smoothly in under 4 seconds at end-expiration; average three values within about 10% of each other. Continuous-output catheters heat blood with a filament and report a rolling average over several minutes, so they lag acute changes.
| Error | Effect on CO reading |
|---|---|
| Less injectate than the computer expects | Falsely high |
| Injectate warmed in the hand | Falsely high |
| Slow or uneven injection | Falsely low |
| Tricuspid regurgitation | Unreliable, usually falsely low |
| Left-to-right shunt (VSD) | Measures pulmonary flow, overstates systemic |
Fick: CO = VO₂ ÷ (CaO₂ − CvO₂) × 10. VO₂ is rarely measured outside the cath lab, and the assumed 125 mL/min/m² can be badly wrong in fever, sepsis or sedation, so indirect Fick is a cross-check. The normal arteriovenous content difference is about 4–6 mL/dL and widens as output falls.
| Feature | SvO₂ | ScvO₂ |
|---|---|---|
| Sample site | PA, distal port | SVC, CVC tip |
| Normal | 60–80% | About 70% |
| Blood included | SVC, IVC and coronary sinus | Upper body only |
| In health | Slightly higher | About 2–3% lower |
| In shock | Lower | About 7% higher, gap widens |
Low SvO₂: delivery is not meeting demand (low output, hemoglobin or SaO₂, or fever, shivering, seizures, pain, work of breathing). High SvO₂ above 80%: tissues are not extracting (sepsis, cyanide from nitroprusside, hypothermia, left-to-right shunt) or a wedged catheter is sampling arterialized blood. A normal ScvO₂ does not exclude hypoperfusion; a venous-to-arterial PCO₂ gap above 6 mmHg suggests flow is still inadequate. ProCESS 2014, ARISE 2014 and ProMISe 2015 found no mortality benefit from ScvO₂-targeted early goal-directed therapy over usual care.
Derived values and worked profiles
| Value | Formula | Normal or threshold |
|---|---|---|
| SVR | (MAP − CVP) × 80 ÷ CO | 800–1200 dyn·s/cm⁵ |
| PVR | (mean PA − PAOP) × 80 ÷ CO | 100–250; above 2 WU is high |
| Diastolic pulmonary gradient | PAD − PAOP | Under 7 mmHg |
| Cardiac power output (CPO) | MAP × CO ÷ 451 | About 1 W at rest; under 0.6 W poor |
| PAPi | (PA systolic − PA diastolic) ÷ RAP | Low values flag RV failure |
| RAP:PAOP ratio | RAP ÷ PAOP | Above about 0.6 suggests RV failure |
CPO is the heart's hydraulic work. In the SHOCK trial registry it was the strongest hemodynamic predictor of in-hospital death, with a cutoff of about 0.53 W (Fincke 2004). PAPi asks how well the RV turns filling pressure into pulsatile PA pressure: under 0.9 predicted RV failure or death in inferior MI (Korabathina 2012), and under about 1.85 predicted RV failure after LVAD (Morine 2016). Thresholds vary by population; treat them as flags.
Worked example: BP 84/52 (MAP 63), HR 110, CVP 14, PA 48/28 (mean 35), PAOP 26, CO 3.2 L/min, BSA 2.0. CI = 1.6 and SV ≈ 29 mL. SVR = (63 − 14) × 80 ÷ 3.2 ≈ 1,225. PVR = (35 − 26) × 80 ÷ 3.2 = 225 (2.8 WU) with a diastolic gradient of 2, so the pulmonary hypertension is post-capillary. CPO = 63 × 3.2 ÷ 451 ≈ 0.45 W; PAPi = 20 ÷ 14 ≈ 1.4; RAP:PAOP ≈ 0.54. Reading: LV-predominant cardiogenic shock with inadequate power and an RV still coping, the profile in which inotropes or LV mechanical support are considered.
| Profile | CVP | PAOP | CI | SVR | SvO₂ |
|---|---|---|---|---|---|
| Hypovolemic | 1 | 4 | 1.8 | 1,600 | 55 |
| Cardiogenic (LV) | 12 | 26 | 1.6 | 1,500 | 48 |
| Septic, after fluid | 6 | 9 | 4.5 | 500 | 80 |
| Massive PE, RV failure | 18 | 8 | 1.7 | 1,500 | 50 |
| Tamponade | 20 | 20 | 1.6 | 1,500 | 50 |
| Neurogenic | 3 | 6 | 2.6 | 600 | 72 |
Minimally invasive monitors, echo and fluid tests
| Monitor | How it works | Main limits |
|---|---|---|
| Uncalibrated pulse contour | Arterial waveform plus demographic model | Changing tone, high-dose pressors, AR, IABP |
| Calibrated pulse contour | Transpulmonary thermodilution resets it | Needs CVC and thermistor arterial line |
| Bioreactance (NICOM) | Phase shift of current across the thorax | Electrodes, motion; mixed agreement |
| Esophageal Doppler | Descending aortic velocity at T5–T6 | Sedated, intubated; probe drifts |
These devices follow trends better than absolute values; uncalibrated pulse contour often misses the 30% percentage-error standard for agreement with thermodilution (Critchley 1999) in vasoplegia and arrhythmia. PPV above 13% predicts response (Michard 2000), but 9–13% is a gray zone holding about a quarter of patients (Cannesson 2011). PPV and SVV also need a heart rate to respiratory rate ratio above about 3.6 and no RV failure, beyond the conditions on the one-pager, and point-prevalence studies (Mahjoub 2014, Delannoy 2016) found few ICU patients meeting all of them. Under 6 mL/kg ventilation, a tidal volume challenge restores some value.
Echo questions a nurse can follow: is the LV squeezing well, poorly or hyperdynamically; is the RV larger than the LV or the septum D-shaped; is there pericardial fluid with RA or RV diastolic collapse? IVC under 2.1 cm collapsing more than 50% on a sniff suggests RAP about 3, over 2.1 cm collapsing less suggests about 15 (ASE 2015); the IVC is unreliable on the ventilator. LVOT VTI is normally about 18–22 cm.
| Test | How | Positive | Limits |
|---|---|---|---|
| Passive leg raise | From 45° semi-recumbent, trunk flat, legs up | CO or SV up 10% within 1–2 min | Needs real-time CO; IAH; high ICP |
| Mini-fluid challenge | 100 mL over 1 min (Muller 2011) | VTI up 10% | Near the noise of echo |
| End-expiratory occlusion | Ventilator hold at end-expiration, 15 s | CO up 5% (Monnet 2009) | No triggering; needs pulse contour |
| Tidal volume challenge | VT 6 to 8 mL/kg for 1 min | PPV up 3.5 points (Myatra 2017) | Controlled ventilation, regular rhythm |
| Fluid challenge | 250–500 mL over 10–15 min | SV up 10–15% | Gives the fluid whatever the answer |
Only about half of ICU patients given fluid raise their output (Michard and Teboul 2002). The leg raise moves about 300 mL of venous blood into the chest and is reversible; the Monnet 2016 meta-analysis gave sensitivity 85% and specificity 91% when judged by output, and much worse by pulse pressure. Pain, coughing or waking during it give false positives. Each test's threshold must exceed its monitor's noise: a 5% rise shows on continuous pulse contour but not reliably on echo. Responsiveness is not an indication; also ask whether low flow is the problem and whether the patient tolerates more volume.
PA catheter complications and evidence
- Insertion: ectopy (common), sustained VT (rare), RBBB, pneumothorax, arterial puncture, air embolism.
- Indwelling: infection, thrombosis, valve injury, knotting; remove when the data no longer change management.
- Balloon: pulmonary infarction from persistent or spontaneous wedge; PA rupture (rare, often fatal); balloon rupture.
| Trial | Population | Result |
|---|---|---|
| Connors 1996 | 5,735 ICU patients, observational | PAC linked to higher 30-day mortality |
| Sandham 2003 | 1,994 high-risk surgical patients | No mortality difference; more PE with PAC |
| PAC-Man 2005 | 1,041 UK ICU patients | Hospital mortality 68% vs 66% |
| ESCAPE 2005 | 433 decompensated HF | Days alive out of hospital 133 vs 135 |
| FACTT 2006 | 1,000 ALI/ARDS patients | No mortality difference; more catheter problems |
| Garan 2020 (CSWG) | Cardiogenic shock registry | Early full profiling linked to lower mortality |
Routine PAC placement does not improve outcomes in general ICU, surgical, ARDS or heart failure patients, and ESCAPE recorded more in-hospital adverse events with it (about 22% vs 12%). It is a monitor, so outcome depends on reading it well; Iberti 1990 found nearly half of surveyed physicians could not identify a wedge on a clear tracing. Use has narrowed to cardiogenic shock (staging, choosing and weaning mechanical support, judging the RV), pulmonary hypertension, transplant and LVAD care, and complex cardiac surgery. Whether PAC-guided care improves survival in cardiogenic shock is unsettled: the supporting data are observational.
Exam pearls
- Transducer below the phlebostatic axis reads high, above reads low; zero to air, level to the 4th intercostal space at mid-chest.
- Square wave: 1.5–2 oscillations; overdamped underreads SBP, underdamped overreads it; MAP is the most trustworthy number.
- Read CVP and PAOP at end-expiration on a frozen trace, at the mean of the a wave.
- PAD exceeding PAOP by more than about 5 mmHg suggests high PVR; PAOP above PAD suggests a non-zone-3 tip.
- Cannon a waves: AV dissociation. Giant v waves on the wedge: mitral regurgitation. Equal CVP, PAD and PAOP: tamponade.
- Too little or warm injectate gives a falsely high thermodilution CO.
- High SvO₂: sepsis, cyanide, hypothermia, wedged catheter.
- CPO under 0.6 W is a poor sign in cardiogenic shock; PAPi flags RV failure.
- Routine PAC does not lower mortality (PAC-Man, ESCAPE, FACTT).
Related one-pagers
- Shock Types at a GlanceHemodynamic patterns, exam findings and first treatments for each shock type
- Vasopressors and InotropesReceptors, doses, SSC 2026 septic shock order, push-dose pressors and extravasation
- ACS and STEMITerritories, time goals, first drugs and complications
- Pericarditis, Effusion and TamponadeDiagnosis, ECG vs STEMI, colchicine and recurrence, tamponade signs and drainage, postoperative and post-MI syndromes
- Aortic Dissection and Hypertensive EmergencyStanford types, heart rate and BP targets, BP goals by condition and IV antihypertensives
- Rhythm Gallery21 drawn strips: how to name each rhythm and the first move exams expect
Calculators for this topic
- Cardiac Index (CI)
- Cardiac Output (Fick Principle)
- Mean Arterial Pressure (MAP)
- Oxygen Delivery (DO₂)
- Pulmonary Vascular Resistance (PVR)
- Stroke Volume (SV)
Sources
- Edwards Lifesciences. Normal Hemodynamic Parameters and Laboratory Values (Edwards Clinical Education pocket card, PP--US-2312 v2.0). 2022. Accessed 2026-10; the file is marked as expiring 2026-04. Link
- Edwards Lifesciences. Normal Hemodynamic Parameters and Laboratory Values (earlier pocket card). 2014. Link
- Edwards Clinical Education. Quick Guide to Cardiopulmonary Care, 4th ed. Edwards Lifesciences. Link
- AACN Practice Alert: Pulmonary Artery/Central Venous Pressure Monitoring in Adults. Crit Care Nurse. 2016;36(4):e12-e18 (reviewed 2021). Link
- Bootsma IT, Boerma EC, de Lange F, Scheeren TWL. The contemporary pulmonary artery catheter. Part 1: placement and waveform analysis. J Clin Monit Comput. 2022;36(1):5-15. Link
- Bootsma IT, Boerma EC, Scheeren TWL, de Lange F. The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications. J Clin Monit Comput. 2022. Link
- Kovacs G, Bartolome S, Denton CP, et al. Definition, classification and diagnosis of pulmonary hypertension (7th World Symposium on Pulmonary Hypertension). Eur Respir J. 2024;64(4):2401324. doi:10.1183/13993003.01324-2024 Link
- Simonneau G, Montani D, Celermajer DS, et al. Haemodynamic definitions and updated clinical classification of pulmonary hypertension (6th WSPH, 2018). Eur Respir J. 2019;53(1):1801913. Link
- CHEST Physician. Major takeaways from the seventh World Symposium on PH. Link
- Prescott HC, Antonelli M, Alhazzani W, et al. Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock 2026. Crit Care Med. 2026. doi:10.1097/CCM.0000000000007075 Link
- Azevedo LCP, Myatra SN, Hammond N. Using the 2026 Surviving Sepsis Campaign Guidelines in Practice. Crit Care Sci. 2026;38:e20260036. Link
- Michard F, Boussat S, Chemla D, et al. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med. 2000;162(1):134-138. Link
- Monnet X, Marik P, Teboul JL. Passive leg raising for predicting fluid responsiveness: a systematic review and meta-analysis. Intensive Care Med. 2016;42(12):1935-1947. Link
- Monnet X, Teboul JL. Passive leg raising: five rules, not a drop of fluid! Crit Care. 2015;19:18. Link
- Delannoy B, Wallet F, Maucort-Boulch D, et al. Applicability of pulse pressure variation during unstable hemodynamic events in the intensive care unit: a five-day prospective multicenter study. Crit Care Res Pract. 2016. Link
- Pearse RM, Hinds CJ. Should we use central venous saturation to guide management in high-risk surgical patients? Crit Care. 2006;10(6):181. Link
- Brittain EL, et al. Predictors of diastolic-to-wedge gradient in patients evaluated for pulmonary hypertension. PLoS One. 2013;8(10):e76461. Link
- Deranged Physiology. Arterial line dynamic response testing (secondary educational source). Link
- Deranged Physiology. Arterial line pressure transducer setup (secondary educational source). Link
- Zhao F, Wei H, Lin L, et al. Levosimendan versus dobutamine in septic cardiomyopathy: a randomized clinical trial. Front Cardiovasc Med. 2025. Link