LVAD Basics

HeartMate 3 parameters, alarms, complications, anticoagulation and arrest care for critical care and transport exams

Reviewed by Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P · Checked against current sources October 2026

Download one-page PDFLetter landscape, ready to print or keep with your notes

The device

Pump parameters

Trend against the patient's baseline; no single number defines status.

ParameterWhat it isWhat changes mean
Speed (rpm)Fixed by the VAD team; adults often 5,000 to 6,000Set speed should not change (except brief PI-event drops)
Flow (L/min)Estimate from speed, power, HctLow: low preload, high MAP, RV failure, tamponade
Power (W)Measured energy to spin the rotorRise: thrombus or aortic insufficiency (report 10 W or more); slow fall: obstruction
PIFlow pulse from native LV beatsDrop: low volume, bleeding, RV failure

PI event (LV collapse sensed): speed drops to the low limit, then ramps back, without an alarm. Thrombus raises power, so calculated flow may read falsely high.

Blood pressure and alarms

AlarmMeaningFirst steps
Low flow (red)Under 2.5 L/minCheck connections; assess volume, bleeding, rhythm, RV
Pump off or driveline out (red)Pump stoppedReconnect; fresh batteries; controller swap if VAD center directs
Low battery (red)Under 5 min leftChange power now
Advisory (yellow)Battery under 15 min or a cable offReconnect or replace power

Complications

ProblemCluesResponse
SuctionLow flow, low PI, PVCs or VT, small LVVolume or blood; treat the cause
Pump thrombosisRising power, high LDH and free Hgb, dark urineIV heparin or bivalirudin; pump exchange if refractory
GI bleedingMelena, falling Hgb (AVMs, acquired vWF loss)Hold or reverse anticoagulation; endoscopy
RV failureHigh CVP, low flow and PIInotropes, inhaled pulmonary vasodilator, RVAD
StrokeNew deficit, severe headacheStat CT/CTA; hold anticoagulation until bleeding excluded
Driveline infectionExit-site redness, drainage, feverCultures, antibiotics; often S. aureus or Pseudomonas

Anticoagulation

Cardiac arrest (AHA 2025)

Check perfusion, not pulse. Adequate if any: normal skin color and temperature, normal capillary refill, MAP over 50 mm Hg (Doppler or arterial line), or PETCO2 over 20 mm Hg via ET tube or tracheostomy (supraglottic airways read falsely high).

  1. Unresponsive: support ventilation, assess perfusion.
  2. Adequate perfusion: no compressions; keep monitoring.
  3. Impaired perfusion: compress now while a second rescuer checks alarms, hum, driveline, power and controller.
  4. Either way, treat non-LVAD causes: arrhythmia, bleeding, low glucose, hypoxia, overdose, RV failure, sepsis, stroke.
  5. Defibrillate and give ACLS drugs as usual; leave the pump connected.
  6. Transport and contact the VAD center.

VT or VF may be tolerated while the pump flows, but treat it; ISHLT advises against letting it run over 30 minutes.

Transport

Practice LVAD Basics questions free in your browser7 exam-style questions with rationales. No account or download needed.

In depth

The one-page summary above is what the PDF prints. These sections go further.

Continuous-flow physiology: why the pulse pressure narrows

A durable LVAD drains the LV apex and returns blood to the ascending aorta, in parallel with the native heart. The rotor runs at one fixed speed, so flow depends on the head pressure across the pump (aortic minus LV pressure): when the native LV contracts, head pressure falls and pump flow rises; in diastole flow falls. That swing is the pulsatility index (PI).

Unloading through the whole cycle lowers LV end-diastolic, wedge and left atrial pressures. Output is pump flow plus whatever the LV still ejects through the aortic valve, which in a well-unloaded ventricle opens intermittently or not at all. Diastolic pressure is held up by continuous flow and systolic pressure is barely augmented, so pulse pressure narrows to roughly 10 to 20 mm Hg: the pulse is weak or absent, oscillometric cuffs and oximeters fail, and the Doppler opening pressure stands in for MAP.

The pump is preload dependent, moving only what reaches the LV, so bleeding, RV failure, tamponade or an arrhythmia that stalls the RV all cut flow, and an emptied ventricle collapses onto the cannula (suction). It is afterload sensitive: at a fixed speed a higher MAP means less flow, and centrifugal pumps have flatter pressure-flow curves than the axial HeartMate II, so hypertension costs them more.

The pump family: HeartMate 3, HeartMate II and HVAD

HeartMate 3 (Abbott) is a centrifugal pump with a fully magnetically levitated rotor, wide flow gaps, textured titanium surfaces and an artificial pulse that Abbott describes as a drop of about 2,000 rpm for 0.15 second, then a rise of 2,000 rpm above set speed for 0.2 second, every 2 seconds, to wash the pump. It sits inside the pericardium, and the modular driveline can be replaced at its connector.

FeatureHeartMate 3HeartMate IIHVAD
Flow designCentrifugal, full magnetic levitationAxial, mechanical bearingsCentrifugal, hydrodynamic plus magnetic
Speed range3,000 to 9,000 rpm; usual 5,000 to 6,0006,000 to 15,000 rpm; usual 8,600 to 9,8001,800 to 4,000 rpm; usual 2,400 to 3,200
Pump positionIntrapericardialPreperitoneal pocketIntrapericardial
Artificial pulseYes, every 2 sNoOptional Lavare cycle
US status 2026Only pump on the marketApproved, being discontinuedWithdrawn June 2021

HeartMate II, an axial pump on mechanical bearings, carried the field from 2008 to 2017 and produced its worst device problem: pump thrombosis at three months rose from 2.2% to 8.4% between 2011 and 2013 (Starling, NEJM 2014). HVAD was withdrawn in June 2021 over failures to restart and a higher stroke rate, leaving several thousand patients on a pump no longer sold. Both still use aspirin and have different controllers and alarms.

Selection, implant and the early postoperative course

Candidates have stage D failure despite guideline therapy: NYHA IIIb to IV, LVEF 25% or less, and inotrope dependence or a peak VO₂ below about 14 mL/kg/min. INTERMACS profiles grade acuity from 1 (critical shock) to 7; most implants are profiles 2 and 3. Destination therapy is roughly 80% of US implants in STS INTERMACS reports since the 2018 allocation change.

The inflow cannula is cored into the apex aimed at the mitral valve, the outflow graft sewn to the ascending aorta and the driveline tunneled to the upper abdomen. The team ramps the speed under transesophageal echo until the septum is midline and the aortic valve opens intermittently; that speed is the baseline you trend against.

The first week is about the right ventricle. Unloading pulls the septum leftward and removes its help to RV contraction while pump flow raises venous return to an RV still facing high pulmonary pressures; it dilates, the LV under-fills, and flow and PI fall with a high CVP. INTERMACS counts right heart failure as needing an RVAD, prolonged inhaled vasodilator or inotropes beyond 14 days, about a quarter of trial patients. Keep CVP in the low teens, add milrinone or dobutamine, use inhaled nitric oxide or epoprostenol, avoid high speed, and move to a temporary RVAD before end organs fail.

Reading the controller and measuring the blood pressure

The screen shows speed, flow, power and PI; only speed and power are measured. Flow is calculated from power and speed with a hematocrit correction and PI from the flow waveform, so whatever distorts power distorts both. No adult normal range is published; the ACTION guide lists flow 3 to 6 L/min, power 3 to 6 W and PI 3 to 5, and each program records the patient's own baseline.

PI is the swing between peak and trough flow against mean flow, averaged over about 15 seconds. A sudden drop against its running average triggers a PI event: speed falls to the team-set low-speed limit so the ventricle can refill, then ramps back without an alarm; each event is logged. A suction event is the clinical version: flow and PI fall, PVCs or VT appear as the cannula irritates the septum, and echo shows a small LV pulled toward the cannula. Restore preload and treat the cause; the team may lower speed, and nobody raises it.

Rotor thrombus adds friction, so power climbs over hours to days or spikes, and because flow is calculated from power the display shows a falsely high flow in an underperfused patient. Hemolysis gives it away: plasma free hemoglobin above about 40 mg/dL, rising LDH, low haptoglobin, tea-colored urine. Aortic insufficiency also raises power and displayed flow, because the regurgitant volume really passes through the pump, but without hemolysis and with a large LV and a jet on echo. Outflow graft obstruction goes the other way: low flow with low or flat power.

Blood pressure is a Doppler skill: manual cuff, Doppler over the brachial or radial artery, inflate until the signal disappears, deflate slowly, and the pressure at which it returns is the opening pressure, which approximates MAP when the pulse pressure is narrow and sits nearer systolic when a pulse is palpable. ISHLT 2023 targets a MAP of 75 to 90 mm Hg (Class IIa) and treats sustained values above 90 with oral antihypertensives, because hypertension cuts flow and predicted stroke in the HVAD trials; Abbott and the ICCAC EMS guide give slightly different ranges.

Alarms, power and the backup controller

HeartMate 3 alarms are two-tiered: red hazard alarms with a continuous tone mean the pump has stopped or is about to; yellow advisories with an intermittent beep need attention within minutes. The silence button quiets a hazard briefly and an advisory for hours, and fixes nothing.

Alarm (color)MeansDo
Pump off (red)Pump stopped: power, driveline or controllerReconnect power or driveline; swap controller if still off
Low flow (red)Under 2.5 L/min (2.0 on low-flow controller)Patient first: volume, bleeding, rhythm, RV, connections
Low battery (red)Under about 5 minutes of powerFresh batteries or MPU now, one source at a time
Controller fault (red)Controller hardware failureSwap to backup controller; call the VAD center
Power cable disconnected (yellow)One of two power leads offReconnect; the other lead is running the pump
Low battery (yellow)Under about 15 minutes of powerReplace batteries before it turns red

The pump needs one live source at all times: pull one battery, replace it, confirm the controller shows power, then change the other; the Mobile Power Unit is the same, one lead at a time. Pulling both stops the pump instantly. A charged pair of 14-volt batteries lasts about 10 to 17 hours.

  1. Controller change (for a failed controller or when the VAD center directs): sit or lie the patient down, because the pump stops briefly.
  2. Connect the backup controller to power first; it alarms until a pump is attached.
  3. Unlock the driveline from the failed controller, seat and lock it in the backup; the pump restarts.
  4. Confirm the hum, the speed and the patient, unplug the old controller, and tell the coordinator how long the pump was stopped.

Complications: bleeding, stroke, infection, RV failure

Bleeding is the most frequent adverse event: in MOMENTUM 3 about 44% of HeartMate 3 patients bled by two years and a quarter bled from the gut. Shear inside the pump unfolds large von Willebrand multimers so ADAMTS13 cleaves them, an acquired von Willebrand syndrome present in nearly every continuous-flow patient, and low pulsatility promotes angiodysplasia in the stomach, duodenum and small bowel.

Hold warfarin (and aspirin in HeartMate II or HVAD patients), reverse only for life-threatening bleeding with a high INR, and scope early. For recurrent bleeding ISHLT 2023 lists a lower INR target, octreotide and thalidomide as options with thin evidence.

ComplicationPreventionRecognize and manage
Ischemic strokeINR in range, MAP under 90, treat infectionCT/CTA; thrombectomy for large-vessel occlusion
Hemorrhagic strokeNo supratherapeutic INR, control MAPReverse warfarin (4F-PCC, vitamin K); neurosurgery
Driveline infectionImmobilize driveline, sterile dressing routineSwab and blood cultures, imaging, long antibiotics
Pump thrombosisINR 2.0 to 3.0, avoid low speed, MAP controlHeparin or bivalirudin; exchange if hemolysis persists
Late RV failureAvoid high speed, manage volume, PDE5 inhibitorDiuretics, home milrinone, reassess transplant
Aortic insufficiencySpeed that opens the valve intermittentlyDiuresis; transcatheter or surgical closure

Driveline infection reaches about one patient in four by two years with no difference between pumps. It starts where movement tugs the velour against skin, and the organisms are skin and water flora: Staphylococcus aureus, coagulase-negative staphylococci, Pseudomonas and enteric gram-negatives. Prevention is mechanical: a stabilizer so the line never moves at the skin, sterile dressing changes per protocol, chlorhexidine, showers only with the kit once the tract matures, no submersion ever. Spreading redness, pus, tract tenderness or fever means cultures, imaging and prompt antibiotics; deep infection means months of therapy, debridement and a higher transplant priority.

Arrhythmias, arrest and transport

Ventricular arrhythmias occur in a third to a half of patients in the first year. Because the LV is drained mechanically, sustained VT or even VF can be tolerated for a time; what fails is the RV, so preload, flow and PI fall and suction follows over minutes to hours. Replace potassium and magnesium, give volume if the LV is small, start amiodarone or lidocaine, and cardiovert or defibrillate normally with pads clear of the pump and the controller connected. ISHLT 2023 allows no more than 30 minutes of VT or VF.

In an arrest the second rescuer looks before touching: hum over the left chest, the screen, any red alarm, both power leads, the driveline seat. Pump off with power intact is a controller or driveline problem; a running pump with no perfusion is upstream (hypovolemia, RV failure, tamponade, PE, tension pneumothorax, hypoxia) or downstream (sepsis, anaphylaxis, overdose, intracranial bleed). Epinephrine, amiodarone and shocks are given as usual.

Daily nursing care, documentation and life after the pump

Three paths follow. Transplant: under the 2018 US allocation a stable LVAD patient sits at status 4, can take 30 days at status 3, and rises to status 2 for device infection, malfunction, thrombosis, intractable arrhythmia or RV failure. Destination therapy: median survival with HeartMate 3 exceeds five years in trial follow-up. Recovery: a few percent, mostly young nonischemic or myocarditis patients, wean; the speed is turned down under echo, and typical explant criteria are an LVEF above 40 to 45%, LVEDD under about 55 mm, wedge under 15 mm Hg and cardiac index above 2.4 at minimal support.

Evidence and controversies

ENDURANCE Supplemental (2018) added a blood pressure protocol with a MAP target of 85 or less to the HVAD trial, cut HVAD strokes, and is the main evidence behind today's MAP targets.

Endpoint for ENDURANCE and MOMENTUM 3: survival free of disabling stroke or reoperation to replace the pump at two years.
TrialComparisonKey result
REMATCH 2001 (n=129)XVE vs medical, DT1-yr survival 52 vs 25%
HeartMate II DT 2009 (n=200)HM II vs XVE2-yr survival 58 vs 24%
ENDURANCE 2017 (n=446)HVAD vs HM II, DT2-yr endpoint 55.0 vs 57.4%; stroke 29.7 vs 12.1%
MOMENTUM 3 2019 (n=1,028)HM3 vs HM II2-yr endpoint 76.9 vs 64.8%; thrombosis 1.4 vs 13.9%
MOMENTUM 3 5-yr 2022HM3 vs HM IISurvival 58.4 vs 43.7%
ARIES-HM3 2023 (n=628)Aspirin vs placebo + warfarinNo more thrombosis; bleeding down about a third

MOMENTUM 3 (Mehra, NEJM 2017 to 2019) randomized 1,028 bridge and destination patients to HeartMate 3 or HeartMate II. At two years survival free of disabling stroke or pump reoperation was 76.9% versus 64.8%, driven by thrombosis (1.4% versus 13.9%), reoperation (2.3% versus 11.3%) and stroke (9.9% versus 19.4%); GI bleeding fell (24.5% versus 30.9%) but stayed common, and two-year survival was similar (about 79% versus 77%). Five-year follow-up (JAMA 2022) separated the curves: 58.4% versus 43.7%. ARIES-HM3 (Mehra, JAMA 2023) randomized 628 HeartMate 3 patients on warfarin to aspirin or placebo; placebo did at least as well on the hemocompatibility composite (about 74% versus 68% event-free) with roughly a third less bleeding, and the FDA label changed in August 2024.

ISHLT 2023 (Saeed, JHLT) is the reference guideline: MAP 75 to 90, warfarin to an INR of 2.0 to 3.0 (it still lists aspirin, predating ARIES), speed for intermittent valve opening, compressions when perfusion is inadequate, and a 30-minute limit on VT or VF.

Unsettled: whether aspirin-free therapy extends to HeartMate II and HVAD; whether the postoperative heparin bridge helps or only bleeds; the exact MAP ceiling; and how to prevent de novo aortic insufficiency.

Exam pearls

Practice free in your browser2,450+ exam-style questions with rationales. No account, no ads. Also free on the App Store and Google Play.

Related one-pagers

Sources

  1. American Heart Association. Highlights of the 2025 American Heart Association Guidelines for CPR and ECC (Left Ventricular Assist Devices section and Figure 8, Adult and Pediatric Durable LVAD Algorithm). October 2025. Link
  2. American Heart Association. Adult and Pediatric Durable Left Ventricular Assist Device Algorithm (2025, updated version), in the 2025 AHA algorithm set hosted by CPR Schoolhouse. Link
  3. American Heart Association. Part 10: Adult and Pediatric Special Circumstances of Resuscitation (Left Ventricular Assist Devices section). 2025 AHA Guidelines for CPR and ECC. Circulation. 2025. Link
  4. Saeed D, et al. The 2023 International Society for Heart and Lung Transplantation Guidelines for Mechanical Circulatory Support: A 10-Year Update. J Heart Lung Transplant. 2023;42(7):e1-e222. Link
  5. Abbott. HeartMate 3 Left Ventricular Assist Device: Pump Parameter Overview (MAT-2007803 v3.0). 2025. Link
  6. International Consortium of Circulatory Assist Clinicians (ICCAC). HeartMate 3 Left Ventricular Assist System Emergency Guide, 2020-2021 (hosted by Northwest Community EMS System). Link
  7. ACTION Learning Network. HeartMate 3 LVAD Algorithms and Care Guide. February 2025. Link
  8. Abbott. Abbott Advances Heart Failure Management with Aspirin-Free Regimen for Patients Receiving the HeartMate 3 Heart Pump. Press release, August 21, 2024. Link
  9. American College of Cardiology. ARIES-HM3 trial summary (Mehra MR, et al. JAMA. 2023;330:2171-2181). Link
  10. U.S. Food and Drug Administration. Recalls Related to the HVAD System. Link
  11. Taylor NP. Abbott's latest HeartMate recall tied to 273 injuries, 14 deaths. MedTech Dive. 2024. Link
  12. BrioHealth Solutions. BrioHealth Solutions Reaches One-Year Milestone and 100th BrioVAD INNOVATE Trial Enrollment. Press release, December 22, 2025. Link
  13. RCEMLearning. Left Ventricular Assist Devices in the ED: Normal parameters of an LVAD. Link
  14. University of Rochester Medical Center Advanced Heart Failure. VAD Low Flow Algorithm. Link