Impella Alarms Algorithm
Suction, position, purge and stopped pump
1 Which alarm?
Suction → step 2Position in ventricle or in aorta → step 3Purge pressure low → step 4Purge pressure high or blocked → step 5Impella stopped → step 6
2 Suction: a filling problem
- Lower the P-level
- Check volume status and RV function (CVP, echo)
- Check position by echo
3 Malposition
- Placement signal ventricular, motor current flat: pump too deep
- Placement signal aortic, motor current flat: pump in the aorta
- Reduce to P-2; the provider repositions under imaging
4 Purge pressure low
- Look for leaks and loose connections
- Purge dextrose too dilute: use a higher concentration
- Replace the cassette if needed
5 Purge pressure high
- Look for kinks in the purge tubing
- Dextrose too concentrated: switch to D5W
- Watch motor current
6 Pump stopped
- Try to restart
- If it will not restart, the provider pulls the pump back into the aorta
Drugs in this algorithm
More devices algorithms